VACUCELL
with vacuum, 200°C

 

 

 

 

 

 

 

 







Laboratory incubators

 

Temperature sensitive, easy decomposable or oxidative materials can be dried very tenderly in VACUCELL vacuum drying ovens, where there is the opportunity of extrusion of air by inert gas. Also complicated components with hardly accessible hollow spaces are drying quick and effectively in VACUCELL ovens.

Volume: 22, 55, 111 liters
Working temperature:
5°C above ambient temperature up to 200° C
Door window
Integrated duct for sensors etc. (Ø 40 mm)
Inert gas connection
Needle valve for fine dosing
Pressure resistant inner chamber
Safety valve-door VENTIFLEX

Interior:
stainless steel, mat. No. 1.4301 (AISI 304)

 The versatile standard line with microprocessor control unit                                    Options
  • 3 programs
  • RS 232 - interface for printer or PC - communication
  • delayed heating start and stop function
  • acoustic alarm
  • time range 99 hours 59 minutes
  • mechanical safety thermostat class 2

    

  • separate PT 100-sensor and temperature indicator
  • communication software for PC (Windows)
  • base box Vacustation
  • chemical resistant vacuum pump
  • separator and an emission condenser
  • vacuum controller

 

... standard line
 The high-tech comfort line with multifunctional microprocessor control unit           Options
  • 6 programs
  • chip card system for individual program storage
  • RS 232 - interface for printer or PC - communication
  • delayed heating start and stop function
  • acoustic alarm
  • time range 0-16 years with 1 min. intervals
  • digital safety thermostat class 3
  • real time
  • programming temperature ramps
  • heating sequences
  • program cycles
  • adjustable ventilation rate 10 to 100%(VENTICELL, STERICELL only)

    

  • base box Vacustation
  • chemical resistant vacuum pump
  • separator and an emission condenser
  • vacuum controller
  • communication software for PC (Windows)
  • RS 485 - interface for PC-communication
  • potential - free alarm contact
  • additional PT - sensor for measuring the temperature of the goods with data output to LCD display or transmission to PC
... comfort line
VACUCELL
Technical data  Model
22
55
111
Interior of stainless steel Material No. 1.4301 volume
width
depth
height
l
mm
mm
mm
55
400
390
350
55
400
390
350
111
540
390
530
Shelves shelves guides-impressions in the chamber side walls
number

max. No.
pieces included

5
2

8
2

9
2
Shelves distance height mm
40
40
40
Useful dim. of shelf width x depth mm.
280 x 236
340 x 296
480 x 386
External Dimensions(including door and handle) width
depth
height
max. mm
max. mm
max. mm
560
490
700
620
550
830
760
640
880
Weight net
gross
kg
kg
65
79
98
114
130
150
Working temperature(beginning of the regulation)   from 5° C above ambient temp. to ° C
200
200
200
Temp. deviations according to DIN 12 880 Teil 2, from working Temperature- Al shelves pressure 5-10 mbar spatially at 100° C
spatially at 200° C
up to ± ° C
up to ± ° C
2
< 5
2
< 6
3
< 7
time variation up to ± ° C
0.4
0.4
0.4
Temp. deviations according to DIN 12 880 Teil 2, from working Temperature- ss shelves pressure 5-10 mbar spatially at 100° C
spatially at 200° C
up to ± ° C
up to ± ° C
10
18
10
23
10
time variation up to ± ° C
0.5
1.0
1.0
Time of rise onto 98% voltage 230V - Al shelves press 5-10 mbar onto temp 100° C
onto temp 200° C
minutes
minutes
60
80
65
85
110
130
Time of rise onto 98% voltage 230V - ss shelves press 5-10 mbar onto temp 100° C
0 heat radiation onto temp
200° C
minutes
minutes
130
170
140
180
70
Heat radiation at 100° C
at 200° C
W
W
150
300
260
520
370
750
Vacuum connection vacuum connection
measuring feedthrough
needle valve for inert gas or air

attainable vacuum
DN mm
DN mm

mm
mbar
16
40

8
5.10-3
16
40

8
5.10-3
16
40

8
5.10-3
Electric parameters-
mains 50/60 Hz
max. input
input in standby by mode
current
nominal Voltage

kW
W
A
V

0.8
5
3.5
230
1.2
5
5.2
230
1.8
5
7.8
230
Max. permissible load of shelves one shelf
totally per unit
kg
kg
20
35
25
45
25
65

Note: All technical data are related to 22°C of ambient temperature and ±10% voltage fluctuations (if not specified otherwise).
Changes in the design and make reserved.

Prev
Index
Next