INCUCELL / INCUCELL V
with natural forced air convection, 70°C/ 99.9°C

 

 

 

 

 

 

 

 

 

 


Laboratory incubators

 

Suitable for safe treatment of microbiological cultures. The INCUCELL line produces no noise and provides very soft air convection within the chamber, the variant INCUCELL V (with a fan) has an advantage of more precise temperature distribution with small deviations. These devices can be used especially in biological and microbiological laboratories, quality tests in pharmacy, cosmetics and testing in veterinary medicine and food processing industry.

Volume: 55, 111, 222, 404, 707 liters
Working temperature:
INCUCELL:
5°C above ambient temperature
up to 70.0° C / 99.9° C
INCUCELL V:
10°C above ambient temperature
up to 70.0° C / 99.9° C
Inner glass door
Interior:
stainless steel, mat. No. 1.4301 (AISI 304)

 The versatile standard line with microprocessor control unit                                    Options
  • 3 programs
  • RS 232 - interface for printer or PC - communication
  • delayed heating start and stop function
  • acoustic alarm
  • time range 99 jours 59 minutes
  • mechanical safety thermostat class 3

    

  • access port Ø 25, 50, 100 mm
  • door lock
  • left door versions
  • seperate PT 100 - sensor and display
  • communication software for PC (windows)
  • HEPA - filter for installation in an air inlet (INCUCELL V only)

 

... standard line
 The high-tech comfort line with multi-functional microprocessor control unit           Options
  • 6 programs
  • chip card system for individual program storage
  • RS 232 - interface for printer or PC - communication
  • delayed heating start and stop function
  • acoustic alarm
  • time range 0-16 years with 1 min. intervals
  • digital safety thermostat class 3
  • real time
  • programming temperature ramps
  • heating sequences
  • program cycles
  • adjustable ventilation rate 10 to 100%(INCUCELL, FRIOCELL, CLIMACELL)

    

  • access port Ø 25, 50, 100 mm
  • door lock
  • left door versions
  • HEPA - filter for installation in air inlet(INCUCELL V only)
  • RS 485 - interface for PC-communication
  • communication software for PC (Windows)
  • timer programmable water protected inner socket
  • potential - free alarm contact
  • additional PT - sensor for measuring the temperature of the goods with data output to LCD display or transmission to PC
... comfort line
INCUCELL, INCUCELL V WITH VENILATOR
Technical data  Model
55
111
222
404
707
Interior of stainless steel Material No. 1.4301 volume
width
depth
height
l
mm
mm
mm
55
400
390
350
111
540
390
530
222
540
540
760
404
540
540
1410
707
940
540
1410
Tray racks
No.
max. No.
pieces included
4
2
7
2
10
2
19
2
19
2
Maximum weight of the load * per tray
inside the oven
max. kg
appr. kg
20
50
20
50
30
70
30
100
50
130
Door    No.
1
1
1
1
2
External Dimensions(including door and handle) width
depth
height **
max. mm
max. mm
max. mm
620
640
680F
760
640
860F
760
790
1100F
760
790
1910R
1160
790
1910R
Package dimensions(three layers carton) width
depth
height (incl. Palette)
mm
mm
mm
710
720
900
850
730
1080
850
730
1320
850
860
2150
1250
910
2150
Mass net
gross
kg
kg
55
61
75
84
100
117
150
165
215
233
Working temperature(beginning of the regulation) standard version
plus version
from 5/10° C above ambient temp. to ° C
70
70
70
70
70
Temperature accuracy according to DIN 12 880 T2, at working Temperature with closed air flap and door space deviation
INCUCELL
INCUCELL V
Appr. ± ° C

< 0.5
< 0.3

< 0.5
< 0.3

< 1
< 0.3

< 1
< 0.8

< 1.5
< 1.5
time variation Appr. ± ° C
< 0.2
< 0.2
< 0.2
< 0.2
< 0.2
Time required to reach 37° C with closed air flap and voltage 230 V INCUCELL
INCUCELL V
Cca min.
49
41
57
51
79
66
41
38
59
51
Heat emission at 37° C   W
30
45
45
60
85
Air exchange speed at 37° C INCUCELL
INCUCELL V
Approx/hr
5
45
5
49
5
24
5
18
5
12
Electricity - mains 50/60 Hz max. power consumption
INCUCELL
INCUCELL V
KW

 
0.3
0.7

 
0.3
0.7

 
0.5
0.7

 
0.9
1.3

 
0.9
1.4
standby W
5
5
5
5
5
Current
INCUCELL
INCUCELL V
A
 
1.4
3
 
1.4
3
 
2
3

3.9
5.6
 
3.9
6.1
nominal voltage V
230
230
230
230
230
*Approx. 50% of the tray area can be filled in a way a uniform air circulation is enabled inside the chamber
** Inclusive Foots and Rollers.

Note: All technical data are related to 22°C of ambient temperature and ±10% voltage swing approx. 50% of the tray area can be filled
in a way a uniform air circulation is enabled inside the chamber.
Changes in the design and make reserved.

Index