CLIMACELL
with forced air convection, cooling
and controlled humidity 99.9° C

 

 

 

 

 

 

 

 

 

Laboratory incubators

The CLIMACELL series was specially developed for applications, in which as far as possible exact and reproducible simulation of various enviromental conditions is important, e.g. stability testing of components, packaging materials, food or chemicals, germination studies, plant cell or tissue cultures, insect cultures.

Volume: 111, 222, 404, 707 liters
Working temperature:
without humidity: 0.0° C up to 99.9° C,
with humidity: 10° C up to 90.0° C
Input medium: drinking water
Controlled humidity: 10% - 90%
Microprocessor controlled humidifying and dehumidifying system
Inner glass door

Interior:
stainless steel, mat. No. 1.4301 (AISI 304)

 The high-tech comfort line with multi-functional microprcessor control unit         Options

    

  • 6 programs
  • chip card system for individual program storage
  • RS 232 - interface for printer or
    PC - communication
  • delayed heating start and stop function
  • acoustic alarm
  • time range 0 - 16 years with 1 min. intervals
  • digital safety thermostat class 3
  • real time
  • programming temperature ramps
  • heating sequences
  • program cycles
  • adjustable ventilation rate 10 to 100% (INCUCELL, FRIOCELL, CLIMACELL)

    

 

     

  • interior lighting
  • access ports Ø 25, 50, 100 mm
  • door lock
  • left door versions (excluded volume 404 liters
  • RS 485-interface for PC - communication
  • timer programmable water protected inner socket
  • automatic light and temperature cycles for day/night simulation
  • exposition light (white day light) 6,000 - 13,000 Lx
  • digitally adjustable light 10 - 100%
  • potential - free alarm contact
  • additional PT - sensor for measuring the temperature of the goods with data output to LCD display or transmission to PC
... comfort line
CLIMACELL
Technical data   Model 111 222 404 707
Inner dimensions
Chamber, stainless steel
volume
width
depth
height
l
mm
mm
mm
111
540
370
530
222
540
520
760
404
540
520
1410
707
940
520
1410
Volume of the steam space   cca l 163 299 524 876
Trays, stainless steel
Min. distance between trays
number max./usual
cm
7/2
7
10/2
7
19/2
7
19/2
7
Number of outer metal doors
Number of inner glass doors
     1
1
1
1
1
1
2
2
Admissible weight of trays
  together inside the oven
    per tray
    
kg
kg
 
50
20
 
70
30
 
100
30
 
130
50
Electricity data
 
Protective system
max. power
consumption
mains
50/60 Hz
W
V
2740
230
IP 20
2740
230
IP 20
3340
230
IP 20
3440
230
IP 20
Temperature data
Working temperature
Temperature accuracy
Variation
 
Deviation
Heating/up time to 37° C from
the ambient temperature
Cooling/down time from 22° C
to 10° C
Recovery time after
1 min. door open
 
 
from 0.0° C
 
at 10° C
at 37° C
 
 
 
 
 
 
at 37° C
at 50° C
 
to ° C
 
±° C
±° C
±° C
min 
 
min 
 
 
min
min
 
99.9
 
<0.5
<0.5
<0.2
20
 
<21 
 
 
4
4
 
99.9
 
<0.5
<0.5
<0.3
25
 
<21 
 
 
4
4
 
99.9
 
<1
<1
<0.3
26
 
<21 
 
 
4
4
 
99.9
 
<1
<1
<0.4
27
 
<21 
 
 
4
4
Relative humidity range % 10 - 95 10 - 95 10 - 95 10 - 95
Heat emission at 37° C W 70 97 123 148
Outer dimensions (incl. door
and handle, feet or rollers)
width max. mm 760 760 1010 1460
depth max. mm 640 790 790 790
height max. mm 1100K 1330K 1910K 1910K
Weight net cca kg 101 132 230 270
gross cca kg 131 169 270 316

All technical data are related to 22°C of ambient temperature and ±10% voltage swing approx. 50% of the tray area can be filled
in a way a uniform air circulation is enabled inside the chamber.
Changes in the design and make reserved.