STERICELL
with forced air convection 250°C

 

 

 

 

 

 

 



 

Laboratory drying ovens

For sterilizing, heating or drying purposes. Due to a patented ventilation system the air within the STERICELL-chamber is ventilated in a regular spiral way. This leads to homogenous temperature profile throughout the chamber and short heating times. A safe process is guaranteed by automatic supervision of the door closing status and a special sterilization program.

Volume: 55, 111, 222, 404 liters
Working temperature:
10°C above ambient temperature up to 250°C
Interior:
stainless steel, mat. No. 1.4301 (AISI 304)

 The versatile standard line with microprocessor control unit                                    Options
  • 3 programs
  • RS 232 - interface for printer or PC - communication
  • delayed heating start and stop function
  • acoustic alarm
  • time range 99 hours 59 minutes
  • mechanical safety thermostat class 3

    

  • door window and interior lighting
  • access ports Ø 25, 50, 100 mm
  • door lock
  • left door versions
  • separate PT 100-sensor and display
  • communication software for PC (Windows)
  • HEPA filter for installation in air inlet
  • possible only by types with volume 404 liters
... standard line
 The high-tech comfort line with multifunctional microprocessor control unit           Options
  • 6 programs
  • chip card system for individual program storage
  • RS 232 - interface for printer or PC - communication
  • delayed heating start and stop function
  • acoustic alarm
  • time range 0-16 years with 1 min. intervals
  • digital safety thermostat class 3
  • real time
  • programming temperature ramps
  • heating sequences
  • program cycles
  • adjustable ventilation rate 10 to 100%(VENTICELL, STERICELL only)

    

  • door window and interior lighting
  • access ports Ø 25, 50, 100 mm
  • door lock
  • left door versions
  • communication software for PC (Windows)
  • RS 485 - interface for PC-communication
  • potential - free alarm contact
  • additional PT - sensor for measuring the temperature of the goods with data output to LCD display or transmission to PC
  • possible only by types with volume 404 liters
... comfort line
STERICELL
Technical data  Model
55
111
222
404
Interior of stainless steel Material No. 1.4301 volume
width
depth
height
ltrs.
mm
mm
mm
55
400
390
350
111
540
390
530
222
540
540
760
404
540
540
1410
Trays racks
No.
max. No.
pieces included
4
2
7
2
10
2
19
2
Maximal weight of the load* per tray
inside the oven
kg
kg
20
50
20
50
30
70
30
100
External Dimensions(including door and handle) width
depth
height**
max. mm
max. mm
max. mm
620
640
680F
760
640
860F
760
790
1100F
760
790
1910R
Package dimensions
(three layers carton)
width
depth
height (incl. palette)
mm
mm
mm
710
720
900
850
730
1080
850
730
1320
850
860
2150
Mass net
gross
kg
kg
55
61
75
84
100
117
150
165
Working temperature(beginning of the regulation) standard version from 10° C above ambient temp. to °C
250
250
250
250
Temperature accuracy according to DIN 12 880 Teil 2, at working Temperature with closed air flap and door space deviation to ° C
+10
-0
+10
-0
+10
-0
+10
-0
time variation to ° C
+3
-0
+3
-0
+3
-0
+3
-0
Time required to reach 250° C with closed air flap and voltage 230 V   minutes
49
53
70
58
Heat emission at 250° C   W
590
760
990
1940
Electricity
- mains 50/60 Hz
max. power input
standby by
current
nominal voltage

KW
W
A
V

1.3
5
5.6
230
1.9
5
8.3
230
1.9
5
8.3
230
3.7
5
5.2; 5.2; 5.2
400/3N
*Approx. 50% of the tray area can be filled in a way a uniform air circulation is enabled inside the chamber.
** Inclusive Foots and Rolls.

Note: All technical data are related to 22°C of ambient temperature and ±10% voltage swing (if not specified).
Changes in the design and make reserved.

Prev